CLiMET
| ||||||||||||||||||||||||||||||||||||||||||||||||||
|
CI-3100 21146 - Ethernet Modbus/TCP-IP Dramatically simplifies installation and validation for remote particle counting sensors.
|
Remote particle counter with internal vacuum pump. Continuously counts particles from a fixed position. Measures the essential two channels - 0.5 and 5 microns at 1 cubic foot per minute. Rugged stainless steel enclosure and small footprint.
The new Ethernet Modbus Optical Particle Transducer ushers in a new era for Continuous Monitoring. CLiMET has successfully transferred the "ease of use" of their portable products to cleanroom monitoring. Requires only power cord and CAT 5e patch cable to connect easily to your company Ethernet.
|
||||||||||||||||||||||||||||||
CI-3100 SPECIFICATIONS:
Optics: Elliptical Mirror (Patent # 5,467,189) Size Sensitivity: 0.5 mM and 5.0 mM Concentration Limit: 1.0 X 106 Particles/Ft3 (35.3 X 106 Particles/M3) Sample Flow Rate: 1.0 Cubic feet per minute Flow Rate Control: Automatic Flow Control Set at Factory Flow Rate Indicator: Mass Flow Cell with LED indicator Display: LED Power, Alarm, Count, Flow Status and Laser Status Network Interface: Ethernet 10/100Base-T (auto sensing), half or full duplex (auto-sensing), MODBUS on TCP/IP; HTTP capable Sample Volume: 1 cubic foot or 1 cubic meter, nominal sample times are 60 seconds or 21 seconds but may vary based on flow and laser status during the sample. Sample data updated once per sample volume continuously; rolling data updated 60 times per volume (nominally every second or 35.3 seconds) Dimensions: 6" x 4" x 11.50 inches (15.24 x 10.16 x 29.21 cm) Weight: 7.25 lbs (3.29 Kg) Power: 100 to 240 VAC, 50/60 Hz, 30W Max Operating Temperature: 32 - 100 deg. F (0 - 37.8 deg. C) Storage Temperature: 32 - 155 deg. F (0 - 68.3 deg. C) Humidity: 0 - 100%, non-condensing Recommended cable: Category 5e unshielded twisted pair patch cable External Exhaust: Ships with 1/2 inch hose fitting (standard)
|
|||||||||||||||||||||||||||||||
|
Mail, fax, email or call your
request for information about Climet Particle Counters by contacting |
|||||||||||||||||||||||||||||||