CLiMET
| ||||||||||||||||||||||||
|
CI-95/CI-95+ Microbiral Air Sampler stainless steel viable sampler
|
Features:
|
||||||||||||||||||
| CI-95 SPECIFICATIONS:
Sample LTD:
CI-95: Aluminum sampling head. Autoclavable @ 134
deg. C for 20 minutes. Enclosure: 304L Stainless Steel Sample Flow Rate: 100 liters per minute (0.10 cubic meter or 3.53 cubic feet per minute) Flow Rate Control: Electronic automatic closed loop Battery: Replaceable 9.6 Volts, nickel metal hydride, 7 hours of continuous operation, 2 - 2.5 hours charge time Display: 4x20 character LCD Unit ID: User selectable; 6 characters long User ID: Up to 10 users; 9 characters long Security: Security to prevent settings changes or stopping a sample in progress. Location ID: Up to 100; 15 characters long Input/Output: RS-232 Audible Alarms: Built-in Alarm Limits: + 10% deviation from 100 LPM flow rate Dimensions: H-W-D: 7.5 x 5.5 x 9.9 inches (19.0 x 14.0 x 25.1 cm) Weight: CI-90: 8.25 lb (3.75 Kg)
with battery Power: Auto Switching: 100-240 VAC, 50/60 Hz Operating Range: 32 - 97 deg. F (0 - 36 deg. C), 10-95% RH, non-condensing Calibration: Factory prime calibration of flow system accuracy recommended every 12 months.
|
|||||||||||||||||||
|
Mail, fax, email or call your
request for information about Climet Particle Counters by contacting |
|||||||||||||||||||